Open original FDA record ↗Recall record
- FDA identifier
- Z-1778-2008
- Company in FDA record
- Stryker Howmedica Osteonics Corp.
- FDA classification date
- Aug 27, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The Xcelerate Patella Reamer Shaft Assembly and the Reamer Adapter do not mate properly. This could result in the removal of excessive bone from the patella leading to a thinner patella; a patella that does not have enough bone for adequate engagement with pegs or patella fracture.
Source facts
- Action
- An IMPORTANT MARKET WITHDRAWAL letter was issued on April 19, 2005 and sent Federal Express to those consignees who received the device. The letter instructed them to examine their inventory to identify instruments, reconcile these instruments on an attached accountability form; fax a copy of this form within 2 days of receipt of the letter; retrieve and return affected product; contact Stryker Customer Service to re-order.
- Initiated
- 2005-04-15
- Root cause
- Process control
- Terminated
- 2008-09-24
- Distribution
- Class II Recall - Worldwide Distribution --- including USA and Canada, United Kingdom.
- Recall number
- Z-1778-2008
- Recall status
- Terminated
- CDRH record ID
- 70289
- FDA classified
- 2008-08-27
- Recall event number
- 47829