Functionality in the report sections "Findings Information" and "Summary of Measured Findings" not functioning properly. The corrected values modified in the report are neither saved, nor printed, nor sent to the information system. The system will still use original values.
Source facts
Action
On 03/27/18 Firm became aware that recall was necessary
Firm initiated recall on 03/27/18 via customer information letter SY007/18/P advising that corrected values modified in the report are not save. Software patch VB20A_HF05 is being released to resolve the matter
Initiated
2018-03-27
Root cause
Software design
Terminated
2020-06-15
Distribution
Nationwide
Recall number
Z-1778-2018
Recall status
Terminated
CDRH record ID
163815
FDA classified
2018-05-09
Recall event number
79888