Open original FDA record ↗Recall record
- FDA identifier
- Z-1780-2013
- Company in FDA record
- Philips Healthcare Inc.
- FDA classification date
- Jul 20, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
IntelliVue Smart-hopping 1.4 GHz Access Point may experience unexpected signal loss of several minutes impacting wireless connectivity
Source facts
- Action
- Philips sent an Urgent Medical Device Correction letter dated July 2013 to affected customers. The letter informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer. Philips will perform a firmware upgrade (version C.00..07 for all affected customers at no cost. Philips is asking customers to follow the Action to be Taken by Customer/User section of the Urgent Medical Device Correction notification/Field Safety Notice. If you need any further information or support concerning this issue, please contact Philips Customer Support at 1-800-722-9377
- Initiated
- 2013-07-08
- Root cause
- Software Design Change
- Terminated
- 2017-05-17
- Distribution
- USA Nationwide Distribution
- Recall number
- Z-1780-2013
- Recall status
- Terminated
- CDRH record ID
- 119750
- FDA classified
- 2013-07-20
- Recall event number
- 65622