The outer box label of the product was labeled with a down revision label which did not include the US labeling requirements, specifically the symbol descriptors and also the "Caution: Federal Law (USA) restricts this device to sale by or on the order of a physician." This product was previously labeled for EU distribution only.
Source facts
Action
Myoscience representatives visited both affected sites on June 14, 2013 and relabeled the product in person. Neither site had used the product and was unaffected by the relabeling. No formal communication was issued.
Initiated
2013-06-21
Root cause
Labeling mix-ups
Terminated
2014-07-15
Distribution
US distribution including CA and ID.
Recall number
Z-1781-2013
Recall status
Terminated
CDRH record ID
119527
FDA classified
2013-07-20
Recall event number
65542