Z-1781-2019: GE Centricity PACS-IW versions 3.6.0 through 3.7.3.9 SP1, SP2, SP3 and 3.7.3 SPa10, Model Numbers: a) 205

Recall record Z-1781-2019 for GE Healthcare, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1781-2019
Company in FDA record
GE Healthcare, LLC
FDA record creation date
Jun 13, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential that one or more image series (i.e., all images within an image set) may be missing from an exam without a warning displayed in the viewer.

Source facts

Action
The firm initiated their medical device correction by letter on 05/2/2019. The letter explained the problem and the safety issue and provided the following safety instructions: 1) Process Root Causes: Investigation determined that the original PACS-IW foundation design did not incorporate adequate risk mitigation for this part of the image ingestion process. DFMEA of the image ingestion wing to wing process was completed as part of the investigation, and no additional areas where further mitigations are required were identified. a) If a discrepancy is identified, attempt to retransmit the exam to PACS. b) If retransmission is unsuccessful, contact a GE Healthcare Service representative for assistance in resolving the impacted exam. c) Urgent cases impacted by this issue should be interpreted by the clinician at the modality. 2) Be aware that if DICOM storage commit is configured and in use, the modality will not be sent a commit notification for images impacted by this issue. 3) Attention should be given to prior cases, acquired before the application of the safety instructions given in this communication, because they may be affected by this issue. For historical data inquiries, please contact a GE Healthcare service representative for assistance in identifying any affected images. The units will be upgraded.
Initiated
2019-05-22
Root cause
Software design
Terminated
2024-09-20
Distribution
Worldwide
Recall number
Z-1781-2019
Recall status
Terminated
CDRH record ID
173306
FDA record created
2019-06-13
Recall event number
83002

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗