Open original FDA record ↗Recall record
- FDA identifier
- Z-1782-2010
- Company in FDA record
- Defibtech LLC
- FDA classification date
- Jun 10, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The AED when used with an affected battery pack, may falsely detect an error condition, cancel charge and not provide therapy
Source facts
- Action
- Defibtech, LLC issued an "Urgent Medical Device Safety Information and Correction" notification dated May 20, 2010 by email and hard copy via US Mail. The notification instructed Consignees on proper identification of affected product and steps to take for product replacement. For further information, contact your distributor or Defibtech Technical Support at techsupport@defibtech.com or call 1-877-453-4507 or 1-203-453-4507, 8:30 AM to 5:00 PM Monday - Friday, EST.
- Initiated
- 2010-05-21
- Root cause
- Device Design
- Terminated
- 2020-09-29
- Distribution
- Worldwide Distribution -- United States, Canada, Australia, Jordan, Spain, Israel, UK, Poland, Italy, Netherlands, Malaysia and Ireland.
- Recall number
- Z-1782-2010
- Recall status
- Terminated
- CDRH record ID
- 90995
- FDA classified
- 2010-06-10
- Recall event number
- 55497