Z-1783-2022: Wisp and Wisp Youth Nasal Mask

Recall record Z-1783-2022 for Philips Respironics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1783-2022
Company in FDA record
Philips Respironics, Inc.
FDA classification date
Sep 30, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

No contraindication/warning regarding magnetic components in CPAP masks for patients with implantable devices or metallic splinters in the eyes.

Source facts

Action
The firm started notifying customers of this recall event via telephone on 9/6/2022. Additionally, customers were mailed an URGENT Medical Device Correction Letter dated 9/9/2022 that communicates that the firm is updating the Contraindications and Warning language for the Amara View Full Face Mask, DreamWisp Nasal Mask, DreamWear Full Face Mask, Wisp and Wisp Youth Nasal Mask, and Therapy Mask 3100 NC/SP due to the presence of magnets that may affect implanted medical devices and metallic splinters in the eye. All units of these masks are subject to this recall action. If a patient, a household member, caregiver, or bedpartner that is in close vicinity to patients using the mask is equipped with one of the medical devices or metallic splinters indicated in the updated Contraindications and Warning language, patients are to stop use of the affected mask and properly dispose of it once an alternative mask without magnets has been obtained after consultation with their healthcare provider. If the patient or others around them does not possess an implanted device or metal splinters included in the updated contraindications language, no action is needed. Users and those in the vicinity of the patient while using the mask must ensure the mask is kept at least six inches from medical implants and metallic splinters in the eye. Customers with any questions are to contact Philips Customer Care Solutions at 1-800-345-6443. The firm issued a press release on 9/6/22 and can be found at the following link: https://www.philips.com/a-w/about/news/archive/standard/news/press/2022/20220906-philips-respironics-alerts-customers-worldwide-of-updated-instructions-and-labeling-of-specific-sleep-therapy-masks-that-contain-magnetic-headgear-clips-due-to-potential-risk-of-serious-injury.html. A second Urgent Medical Device Correction Letter was sent to customers on 12/1/2022. Should DMEs, distributors, and prescribers decide to continue distributing masks and mask accessories with magnet
Initiated
2022-09-06
Root cause
Error in labeling
Distribution
Worldwide distribution - US Nationwide distribution including in the states of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West Virginia, Wisconsin, Wyoming, & District of Columbia. The countries of Algeria, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bermuda, Bolivia, Bosnia-Herz., Brazil, Brunei Darussal, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Curacao, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Estonia, Faroe Islands, Finland, France, French Guiana, French Polynesia, Germany, Ghana, Guadeloupe, Guam, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kenya, Korea, Kuwait, Lebanon, Lithuania, Macedonia, Malaysia, Malta, Martinique, Mexico, Montenegro, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Paraguay, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, St. Pier, Miquel., Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Kingdom, United States, Uruguay, Utd. Arab. Emir., Uzbekistan, & Zimbabwe.
Recall number
Z-1783-2022
Recall status
Open, Classified
CDRH record ID
195532
FDA classified
2022-09-30
Recall event number
90705

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗