Open original FDA record ↗Recall record
- FDA identifier
- Z-1784-2012
- Company in FDA record
- Captiva Spine, Inc
- FDA classification date
- Jun 7, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Captiva Spine, Inc. initiated a recall of FuseLOX Lumbar Implants, Trials, and Inserters after FDA advised the firm that there is inadequate support for implemented design changes to the cleared device Pivotec Lumbar Interbody Fusion Device.
Source facts
- Action
- Captiva Spine sent a Voluntary Correction Notification letter dated May 4, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to locate and return affected devices. Customers were also instructed to complete and return a Tracking/Verification Form. For questions customers were instructed to call 877-772-5571. For questions regarding this recall call 561-277-9480.
- Initiated
- 2012-04-27
- Root cause
- Finished device change control
- Terminated
- 2013-12-24
- Distribution
- USA (nationwide) including FL,TX, UT, and NJ.
- Recall number
- Z-1784-2012
- Recall status
- Terminated
- CDRH record ID
- 109498
- FDA classified
- 2012-06-07
- Recall event number
- 61925