Open original FDA record ↗Recall record
- FDA identifier
- Z-1784-2024
- Company in FDA record
- NRT X-RAY A/S
- FDA classification date
- May 7, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
X-ray system C-arm may experience uncontrolled motion if 1) AC motor controller firmware has a faulty setting, 2) the Motor/gear assembly is worn, 3) CAT movement is driven at maximum speed and angulated more than 75 deg. from vertical, and 4) the emergency stop or touch guard is activated during the breaking sequence, which could lead to the C-arm hitting the patient or operator.
Source facts
- Action
- On 04/17/24, correction notices were emailed to consignees who were asked to do the following: 1) If you experience the described uncontrolled motion, stop using the device and contact your service provider to have your device repaired. 2) The correction notice should be shared with anyone who needs to be aware within your organization and forwarded to any organization where potentially affected devices have been transferred. 3) Complete and return the response form to either support@nrtxray.com or servicesupport@alpha-imaging.com A firmware update will be made to all motor controllers in the system. The update will be performed by the Service Provider. Customers with recall-related questions can call 1-800-331-7327.
- Initiated
- 2024-04-17
- Root cause
- Device Design
- Distribution
- US Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, WV, LA, NY, PA, HI, NV, WI, MN, MI.
- Recall number
- Z-1784-2024
- Recall status
- Open, Classified
- CDRH record ID
- 207034
- FDA classified
- 2024-05-07
- Recall event number
- 94363