Public FDA record

Z-1784-2025: DxC 500 AU Clinical Chemistry Analyzer, Catalog Numbers/UDI codes: C63519 / 14987666545058 C63520 / 14987

Recall record Z-1784-2025 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1784-2025
Company in FDA record
Beckman Coulter Inc.
FDA classification date
May 15, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Beckman Coulter is recalling their DxC 500 AU Clinical Chemistry Analyzer because a software error causes the analyzer to not run a requested calibration order in the following scenario: when a reagent blank or calibration is ordered during sample processing and then any of the components (R1 and/or R2) depletes to zero tests, the analyzer will not be able to complete the calibration request, and the calibration order will remain pending. No further calibration orders can be processed for any assays, and the instrument refuses to accept further sample processing order after the existing calibration curves are expired. Although in-process tests will be completed, this error can cause a delay in reporting subsequent test results. No further calibration orders can be processed for any assays which may cause a delay in reporting test results.

Source facts

Action
On or about 03/20/2025, the firm sent via email or postal mail an Urgent Medical Device Recall letters informing customers that Beckman Coulter has determined that the DxC 500 AU Clinical Chemistry Analyzer may not run a requested calibration order for a two-part reagent and subsequently prevents any new calibration order from being placed. For two-part reagents, when reagent blank or calibration is ordered during sample processing and then any of the components (R1 and/or R2) depletes to zero tests, the analyzer will not be able to complete the calibration request, and the calibration order will remain pending. The analyzer will not allow additional calibration orders to be requested for any assays. Customers are instructed to: . Do not place calibration orders when the analyzer is processing samples. . Refer to the DxC 500 AU IFU Reagent Overview section for instructions on Monitoring the Status of Reagents. . Beckman Coulter recommends sharing the content of this letter with their laboratory and/or Medical Director. . If they experience this issue on software version 1.3, 1.4, 1.4.1 or 1.4.2, contact Beckman Coulter for service. . Beckman Coulter recommends posting this Recall Letter on or near the affected systems until the corrections have been made. Resolution: . This has been resolved in software version 1.4.3 in December of 2024. . Your Beckman Coulter service representative will contact you to schedule the software upgrade. For questions - contact Customer Support Center at: . From our website: http://www.beckmancoulter.com . Troubleshooting Hotline: (800) 854-3633
Initiated
2025-03-20
Root cause
Software design
Distribution
U.S: AL, CA, GA, ID, IN, LA, MA, MI, MN, MT, NC, NV, PA, PR, TN , TX, and WI O.U.S.: Australia, Belgium, Bosnia and Herzegovina, Canada, Chile, Croatia, Egypt, Ethiopia, France, Germany, Ghana, India, Italy, Korea, Lebanon, Libya, Malaysia, Netherlands, North Macedonia, Oman, Paraguay, Philippines, Poland, Puerto Rico, Qatar, Saudi Arabia, South Africa, Spain, Switzerland, and Tunisia
Recall number
Z-1784-2025
Recall status
Open, Classified
CDRH record ID
213083
FDA classified
2025-05-15
Recall event number
96512

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗