Z-1785-2019: Liquid Cardiac Control, Model Nos. CQ5051, CQ5052, CQ5053

Recall record Z-1785-2019 for Randox Laboratories Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1785-2019
Company in FDA record
Randox Laboratories Ltd.
FDA record creation date
Jun 14, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Customers may observe a decrease in recovered concentration with regards to NTproBNP in the Liquid Cardiac Control lots identified, on the Siemens Dimension EXL LOCI.

Source facts

Action
On March 11, 2019, the firm notified affected consignees of the product issue through an Urgent Medical Device Correction letter. Customers were informed that they may observe a decrease in recovered concentration for NTproBNP in the referenced Liquid Cardiac Control lots compared to the quoted target value on the value sheet for this analyzer only (Siemens Dimension EXL LOCI). Customers were asked to do the following: * Inspect your stock and quarantine affected stock. " Replace the value sheet in the kit with the revised value sheet provided. " Randox is not recommending a review of previous results as changes in quality control recovery would be reviewed at the time of occurrence. " Discuss the contents of this notice with your Medical Director. " Complete the response form even if you no longer have the affected product. Return the response form to technical.services@randox.com within five working days.
Initiated
2019-03-11
Root cause
Under Investigation by firm
Terminated
2020-05-15
Distribution
Product distributed to CA, DE, WV, and Puerto Rico.
Recall number
Z-1785-2019
Recall status
Terminated
CDRH record ID
171548
FDA record created
2019-06-14
Recall event number
82400

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗