Skip to content

Public FDA record

Z-1785-2025: True 3D Viewer, Model: EP-00003

Recall record Z-1785-2025 for Echopixel, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1785-2025
Company in FDA record
Echopixel, Inc.
FDA classification date
May 16, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

With certain software versions, import of enhanced ultrasound formatted data/ultrasound images into 3D viewer, (for digital image processing/review/analysis/communication/media interchange acquired from CT/MRI/ultrasound, and pre-operative analysis of surgical options), can result in incorrect display of image orientation, which could result in misdiagnosis or incorrect surgical treatment plan.

Source facts

Action
On 2/21/2018, correction notices were emailed to customers who were asked to do the following: a) Affected software versions should not be used with Enhanced 3D Ultrasound DICOM compliant volumes or Philips 3DDCM. b) If clinicians are utilizing Enhanced 3D Ultrasound DICOM compliant volumes or Philips 3DDCM volume data sets, verify, retrospectively, any potential risk to health to patients and further determine if patients need to be reevaluated and report this information to the firm. c) Pass a copy of this notice to those staff members who need to be aware of the issue and maintain awareness for a three-month period. d)If devices were transferred to other departments or organizations, provide a copy of this notice to these consignees. Firm will contact customers to provide information about the timing and delivery mechanisms of the software upgrade. If you have questions, contact the firm at: 1-844-273-7766 x2
Initiated
2018-02-21
Root cause
Device Design
Distribution
Worldwide - US Nationwide distribution in the states of CA, OH, NJ, PA, MA, FL, CT, TX, LA, MI, MN and the country of CA.
Recall number
Z-1785-2025
Recall status
Open, Classified
CDRH record ID
213387
FDA classified
2025-05-16
Recall event number
96674

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗