Skip to content

Public FDA record

Z-1786-2010: directCHECK Whole Blood Control; for Hemochron J. Microcoagulation Systems.

Recall record Z-1786-2010 for International Technidyne Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1786-2010
Company in FDA record
International Technidyne Corporation
FDA classification date
Jun 11, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Indication not cleared or approved. The acceptable performance range published in the directCHECK Microcoagulation Quality Control package insert is incorrect.

Source facts

Action
Notification letters were sent vis Federal Express on April 15, 2010 to all distributors and end users to return unused product or to replace the package insert. Please direct questions ITC Technical Support at phone 800-631-5945 Ext. 4707.
Initiated
2010-04-15
Root cause
Error in labeling
Terminated
2010-09-23
Distribution
Distributors and medical facilities in AZ, CA, IL, KY, MA, MD, ME, MO, PA, WA. Foreign distribution to Australia, Canada, Switzerland, Germany, Spain, Quatar, and Saudi Arabia.
Recall number
Z-1786-2010
Recall status
Terminated
CDRH record ID
90767
FDA classified
2010-06-11
Recall event number
55404

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗