A bearing used in the assembly of the Fine Traction Device did not perform successfully in pre-release reliability testing; a replacement bushing passed reliability testing and was substituted per a change order. The change order authorized use of the substitute bushing in assembly of the Fine Traction Accessory; however, inadvertently the new part was not implemented. During complaint evaluati
Source facts
Action
STERIS Account Managers notified affected customers via onsite visits. All consignees were notified as of 5/29/2013. Customers were instructed to return affected product for a replacement unit.
Initiated
2013-05-20
Root cause
Component change control
Terminated
2013-08-28
Distribution
USA Nationwide Distribution in the states of CO, KY, LA, VA, and WI.
The affected product was also distributed to Canada.
Recall number
Z-1786-2013
Recall status
Terminated
CDRH record ID
119612
FDA classified
2013-07-22
Recall event number
65573