Z-1786-2022: MICRO-X ROVER MOBILE X-RAY SYSTEM, MXU-RV19

Recall record Z-1786-2022 for Micro-X Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1786-2022
Company in FDA record
Micro-X Ltd.
FDA classification date
Sep 23, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Software calibration error with product equip with a Dose Area Product (DAP) meter. This results in the product displaying the incorrect DAP meter values which may impede radiation exposure management decisions

Source facts

Action
On or about 09/31/2022, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Letter via email/telephone to customers informing them that during an internal data analysis on 08/18/2022, Micro-X Ltd identified software calibration errors with the product equipped with a Dose Area Product (DAP) meter. This resulted in the product displaying incorrect DAP meter values. The DICOM image produced by the system will report a DAP measurement in microgray/square meter that is 8 to 10 times lower than the actual dose delivered. Customers are instructed to: 1. Inspect their stock immediately and place the Recall Notification Letter on the affected product and quarantine affected product on hand to prevent further use. 2. Contact the nominated Service representative to arrange the re-calibration of the DAP Meter. Email: support-us@micro-x.com or Phone 1-206-249-8764. 3.Previous version of Service Documentation (ROV208-2 Acceptance and Compliance Testing Instructions) which includes inaccurate DAP calibration processes are to be removed from use. 4. Complete and return Acknowledgement and Receipt Form immediately to Micro-X LTD in Australia: support@micro-x.com and copy djerkovic@micro-x.com 5. Ensure relevant staff members are informed of the Recall and if affected units have been supplied/transferred to another facility, inform them immediately. For question or further assistance, contact Regulatory Affairs Manager at djerkovic@micro-x.com
Initiated
2022-08-31
Root cause
Process control
Distribution
Worldwide distribution - U.S. Nationwide distribution in the states of CA, NV, NY, OH, TX and WA and the countries of Australia, France, Israel, Japan, and United Arab Emirates.
Recall number
Z-1786-2022
Recall status
Open, Classified
CDRH record ID
195990
FDA classified
2022-09-23
Recall event number
90839

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗