Open original FDA record ↗Recall record
- FDA identifier
- Z-1787-2012
- Company in FDA record
- GE Healthcare, LLC
- FDA classification date
- Jun 17, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class I
Reason stated in the source
GE Healthcare has recently become aware of a potential safety issue due to the interlock slide mechanism missing from the 3-vaporizer manifold of your Aestiva 7900 Anesthesia Device. This Potential for two (2) vaporizers delivering agent at the same time. This could manifest as over delivery of a single agent if both vaporizers contain the same agent or as delivery of more than one agent. This m
Source facts
- Action
- GE Healthcare sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated April 10, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. Contact GE Healthcare Customer or Technical Support at 800-345-2700 for questions regarding this notification.
- Initiated
- 2012-04-10
- Root cause
- Process control
- Terminated
- 2012-07-09
- Distribution
- Nationwide Distribution - including the states of AL, FL, GA, MO, and VA.
- Recall number
- Z-1787-2012
- Recall status
- Terminated
- CDRH record ID
- 108604
- FDA classified
- 2012-06-17
- Recall event number
- 61639