Z-1788-2024: RKN130003, RHEO KNEE 3 YR WARRANTY, MD, CE, External lower limb prosthetic component

Recall record Z-1788-2024 for Ossur H / F. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1788-2024
Company in FDA record
Ossur H / F
FDA classification date
May 10, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to firmware issues with the prosthetic knee, there is the potential for unintended warnings and device shutdown which could result in patient falls.

Source facts

Action
On April 4, 2024, Ossur, issued a Urgent Medical Device Correction via E-Mail to affected consignees. Ossur asked consignees to take the following actions: REMOTELY UPDATE FIRMWARE ON AFFECTED DEVICES 1a. Check the serial number of devices against the list of affected devices in this letter to confirm the affected units. 2b. Update the firmware on the device via the Rheo Knee Update application This application can only be used to update the affected devices: i. A computer running Windows Operating System 10 or 11 is required to run the Rheo Knee Update application. It is not compatible with Android or iOS and cannot be completed via mobile phones. ii. Access the Rheo Knee Update application via this link: htps://www.ossur.com/global/prismic/generic/rheo-knee-update-language iii. Follow the steps outlined in the application to complete the firmware update. iv. If there are any issues with completing the firmware update successfully, please contact ssur technical support at 1-800-233-6263 ext. 10. 2. PLEASE PASS THIS NOTICE to those who need to be aware within your organization. If you have further distributed this product, please identify your customers, and notify them at once of this product alert. We recommend that you include a copy of this notice. 3. PLEASE MAINTAIN AWARENESS of this notice and required actions for an appropriate period to ensure effectiveness of the corrective action.
Initiated
2024-03-25
Root cause
Software design
Distribution
US Nationwide distribution in the states of AL, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NH, NJ, NM, NY, OH, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY.
Recall number
Z-1788-2024
Recall status
Open, Classified
CDRH record ID
206946
FDA classified
2024-05-10
Recall event number
94332

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗