Skip to content

Public FDA record

Z-1790-2008: Psi-Tec Tubing, 12'; 10/box; Part Number PT-5558; distributed by Mentor Texas, LP, Irving, TX 75038.

Recall record Z-1790-2008 for Mentor Texas, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1790-2008
Company in FDA record
Mentor Texas, Inc
FDA classification date
Sep 5, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

During packaging integrity validation, it was found that various product packages could fail, compromising the sterility of the devices.

Source facts

Action
A Customer Letter was issued to all consignees by the firm on 4/21/2008. Consignees were informed that a recent testing of these products discovered a potential defect in the packaging configuration, which could result in a loss of package integrity. Consignees were asked to review inventory and dispose of product and contact firm for credit.
Initiated
2008-04-21
Root cause
Other
Terminated
2010-07-29
Distribution
Class II Recall - Worldwide Distribution ---- including USA, Puerto Rico and countries of Argentina, Austria, Bermuda, Brazil, Canada, China, Costa Rica, Cyprus, Denmark, England, Germany, Greece, Guatemala, India, Italy, Italy, Japan, Kuwait, Lebanon, Malta, Mexico, Morocco, Pakistan, Panama, Paraguay, Portugal, Slovak Republic, Slovenia, Spain, South Africa, South Korea, Switzerland, Taiwan, Thailand, Turkey and Venezuela.
Recall number
Z-1790-2008
Recall status
Terminated
CDRH record ID
70397
FDA classified
2008-09-05
Recall event number
47873

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗