- FDA identifier
- Z-1790-2020
- Company in FDA record
- CSL Behring GmbH
- FDA record creation date
- Apr 23, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Negative bias for the carbohydrate-deficient transferrin (CDT) measurement in comparison to the HPLC method, observed to occur after 15 months based on shelf life monitoring
Source facts
- Action
- Siemens Healthcare Diagnostics issued Urgent Medical Device Correction (UMDC) letter PP-20-002.A.US to the 5 affected customers in the U.S. on March 5, 2020 via FedEx overnight delivery. ( Urgent Field Corrective Action (UFCA) PP20-002.A.OUS was issued to all OUS Siemens sites with affected customers electronically for distribution and regional reporting to their local competent authorities). Letter states reason for recall, health risk, and action to take by the Customer Please review this letter with your Medical Director. Discontinue use of and discard the kit lots listed in Table 1. Review your inventory of these products to determine your laboratorys replacement needs and to provide information to Siemens Healthineers for reporting to the authorities. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.
- Initiated
- 2020-03-02
- Root cause
- Under Investigation by firm
- Terminated
- 2020-10-05
- Distribution
- IN, KS, MO, OH, VA Foreign: Australia, Austria, Bangladesh, Belgium, Canada, China Czech Republic, Egypt, Finland, France, Germany, Great Britain, Hungary India, Italy, Japan, Mexico, Netherlands, Norway, Poland, Portugal, Republic Korea, Russia Federation, Slovakia, Slovenia, South Africa, Spain Sweden, Switzerland, Taiwan
- Recall number
- Z-1790-2020
- Recall status
- Terminated
- CDRH record ID
- 180413
- FDA record created
- 2020-04-23
- Recall event number
- 85282