Skip to content

Public FDA record

Z-1790-2021: The Sarns Temperature Control and Monitor unit (TCM) is a source of temperature-controlled water for bloo

Recall record Z-1790-2021 for Terumo Cardiovascular Systems Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1790-2021
Company in FDA record
Terumo Cardiovascular Systems Corporation
FDA record creation date
Jun 4, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Terumo CVS has been unable to validate a cleaning protocol to satisfy current regulatory concerns and expectations. As a result, an updated cleaning protocol will not be developed by Terumo CVS and it has been determined that the best course of action is for users to discontinue use of and dispose of HX2, TCM I and TCM II devices.

Source facts

Action
On 04/30/2021, Terumo issued an Urgent Medical Device Removal notice to customer via letter notifying users to discontinue the use of and dispose of HX2, TCM I and TCM II devices. Customers were instructed to confirm receipt of this communication by completing and returning the attached Customer Response Form as indicated on the form. For questions contact Terumo CVS Customer Service: 1-800-521-2818.
Initiated
2021-04-16
Root cause
Device Design
Terminated
2023-11-29
Distribution
Domestic: Foreign: Australia, Belgium, Canada, Chile, China, Colombia, Dominican Republic, England, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Israel, Italy, Japan, Korea, Malaysia, Mexico, Netherlands, New Zealand, Norway, Philippines, Russia, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UNITED ARAB EMIRATES (UAE), Vietnam
Recall number
Z-1790-2021
Recall status
Terminated
CDRH record ID
187215
FDA record created
2021-06-04
Recall event number
87834

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗