Open original FDA record ↗Recall record
- FDA identifier
- Z-1792-2024
- Company in FDA record
- TriMed Inc.
- FDA classification date
- May 10, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Nail system contains an out of specification driver-socket which could result in inability to pick up nails or apply sufficient torque.
Source facts
- Action
- On 4/3/24, TriMed Inc., issued a "Urgent: Medical Device Recall", to affected consignees via E-Mail. TriMed ask consignees to take the following actions: " If you have the affected product, discontinue further use and quarantine products prior to return. " If a successful implantation was completed no further action is necessary. Only standard monitoring of the patient is required after successful implantation. " Please share this notice with all personnel that may utilize the affected products. Please share this notification to any locations where the affected product may have been transferred to. " Conduct a physical count of the affected product in your possession and record the count on attached response form. Once the response form is completed, email to: Quality@trimedortho.com. Please respond even if you have no affected products in your inventory. " Once the response form is completed and emailed, you will receive an RGA (Return Good Authorization form) as well as a link to generate a label to return the devices via email. Please package the devices, print the return label and ship via FedEx.
- Initiated
- 2024-03-19
- Root cause
- Process control
- Distribution
- US Nationwide distribution in the states of AK, AR, AZ, CA, FL, HI, IL, LA, MA, MI, NC, NJ, NV, OH, OK, PA, TN, TX, UT, VA, WA, WI.
- Recall number
- Z-1792-2024
- Recall status
- Open, Classified
- CDRH record ID
- 207099
- FDA classified
- 2024-05-10
- Recall event number
- 94381