Skip to content

Public FDA record

Z-1793-2020: PREVI Isola Insrument and PREVI Isola Printer spares PSUs (power supply units), P1028888-001 PSU (used i

Recall record Z-1793-2020 for bioMerieux, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1793-2020
Company in FDA record
bioMerieux, Inc.
FDA record creation date
Apr 23, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The recall has been initiated due to potential for overheating or a fire hazard associated with the Zebra LP2824 Plus printer PSU (power supply unit) located within the PREVI Isola instrument.

Source facts

Action
The firm initiated the recall by letter on 01/15/2020. The Urgent Product Correction Notice instructed the customer the to provide the letter to all appropriate personnel within their organization. The letter instructs the personnel to check their inventory against the serial numbers listed in Appendix A of the letter. The letter also details that a bioMerieux Field Service Engineer will be contacting the customer to schedule a service appointment and if necessary, replace the PSU.
Initiated
2020-01-24
Root cause
Nonconforming Material/Component
Terminated
2023-01-11
Distribution
Worldwide distribution. US nationwide including AU, BE, CH, DE, ES, FR, IT, HK/CN; MX, NL, RU, SE/FI, ZA.
Recall number
Z-1793-2020
Recall status
Terminated
CDRH record ID
179235
FDA record created
2020-04-23
Recall event number
84737

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗