Open original FDA record ↗Recall record
- FDA identifier
- Z-1793-2026
- Company in FDA record
- Spacelabs Healthcare, Inc.
- FDA classification date
- Apr 7, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO measurement not automatically excluded, which may result in inaccurate readings, or the need of additional readings, which may lead to delayed care and/or fluid overload.
Source facts
- Action
- On 2/27/2026, correction notices were emailed to customers who were asked to do the following: 1. Notify all relevant employees within your organization. If products were further distributed, please forward the notice to these sites/customers. 2. Complete and return the acknowledgement form via email to: Installedbase@Spacelabs.com 3. Instruct staff who utilize the cardiac output feature to use the "Manual" mode, or use a non-impacted module in its place (if available) until the impacted module has been corrected. 4. Contact your local Spacelabs Healthcare service and support representative to schedule your device correction as soon as possible. Firm's Support Contact: Phone: 1-800-522-7025 Email: Support@Spacelabs.com Website: https://spacelabshealthcare.com/support/
- Initiated
- 2026-02-27
- Root cause
- Component change control
- Distribution
- Worldwide - US Nationwide distribution including in the states of NY, NJ, AL, OK, IA, MT, WY, MN, TX, WI, PA, WA, OR, MI, CA, SC, ID, FL, MS, KS, SD, LA, OH, MO, NH, GA, MD, TN, AZ, PR and the countries of TAIWAN, CANADA, KUWAIT, FRANCE, SAUDI ARABIA, POLAND, PAKISTAN, MEXICO, ARGENTINA, PANAMA.
- Recall number
- Z-1793-2026
- Recall status
- Open, Classified
- CDRH record ID
- 218485
- FDA classified
- 2026-04-07
- Recall event number
- 98374