Skip to content

Public FDA record

Z-1793-2026: Multi-parameter Command Module, Model 91496, optioned with Masimo or Nellcor

Recall record Z-1793-2026 for Spacelabs Healthcare, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1793-2026
Company in FDA record
Spacelabs Healthcare, Inc.
FDA classification date
Apr 7, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO measurement not automatically excluded, which may result in inaccurate readings, or the need of additional readings, which may lead to delayed care and/or fluid overload.

Source facts

Action
On 2/27/2026, correction notices were emailed to customers who were asked to do the following: 1. Notify all relevant employees within your organization. If products were further distributed, please forward the notice to these sites/customers. 2. Complete and return the acknowledgement form via email to: Installedbase@Spacelabs.com 3. Instruct staff who utilize the cardiac output feature to use the "Manual" mode, or use a non-impacted module in its place (if available) until the impacted module has been corrected. 4. Contact your local Spacelabs Healthcare service and support representative to schedule your device correction as soon as possible. Firm's Support Contact: Phone: 1-800-522-7025 Email: Support@Spacelabs.com Website: https://spacelabshealthcare.com/support/
Initiated
2026-02-27
Root cause
Component change control
Distribution
Worldwide - US Nationwide distribution including in the states of NY, NJ, AL, OK, IA, MT, WY, MN, TX, WI, PA, WA, OR, MI, CA, SC, ID, FL, MS, KS, SD, LA, OH, MO, NH, GA, MD, TN, AZ, PR and the countries of TAIWAN, CANADA, KUWAIT, FRANCE, SAUDI ARABIA, POLAND, PAKISTAN, MEXICO, ARGENTINA, PANAMA.
Recall number
Z-1793-2026
Recall status
Open, Classified
CDRH record ID
218485
FDA classified
2026-04-07
Recall event number
98374

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.