Z-1794-2026: MicroScan Neg Multidrug Resistant MIC 1 REF C32368 UDI-DI code: 15099590720568 Intended Use- For use wit

Recall record Z-1794-2026 for Beckman Coulter, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1794-2026
Company in FDA record
Beckman Coulter, Inc.
FDA classification date
Apr 8, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to manufacturing issue (incorrectly processing) with in vitro multidrug resistant test.

Source facts

Action
On 09/08/2025, the firm sent via mail/email an "URGENT FIELD SAFETY NOTICE" to customers informing them that as a result of an internal investigation Beckman Coulter confirmed that Ciprofloxacin (Cp) antimicrobial agent was incorrectly manufactured on MicroScan Neg Multidrug Resistant MIC 1 Lot 2026-05-08. Customers are instructed to: "Review Ciprofloxacin results generated from Lot 2026-05-08 to identify any abnormal or unexpected trends in both susceptible and resistance patterns, or unexpected change in MIC trends, as compared to historical results for that same species (e.g. a sudden, unexplained increase in susceptible or % susceptible results for a species after implementation of Lot 2026-05-08 versus the previous susceptible or % susceptible results for the same species prior to using Lot 2026-05-08). "Discontinue use of this lot and discard per your laboratory guidelines "The laboratory should retain the inventory of other MicroScan Neg Multidrug Resistant MIC 1 lots as they are not impacted by this issue. Note-Beckman Coulter is no longer distributing MicroScan Neg Multidrug Resistant MIC 1 Lot 2026-05-08 For questions - contact Customer Support Center from website: http://www.beckmancoulter.com
Initiated
2025-09-08
Root cause
Under Investigation by firm
Distribution
International distribution to the country of Poland.
Recall number
Z-1794-2026
Recall status
Open, Classified
CDRH record ID
219085
FDA classified
2026-04-08
Recall event number
98562

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗