Siemens issued a Field Safety Notice about the potential hazard to patients or operators when using the function "table top float" of the AXIOM Aristos MX. When repositioning a patient lying on the table top and releasing the table top, it is possible that the fingers of the patient or operator are between the table top and the detector tray and if the table top is moved, the fingers can be squee
Source facts
Action
On 6/5/13, consignees were notified by letter of a "Field Safety Notice" and received instructions about how to avoid the potential risk of this issue while using the device. Siemens also rolled out a modification of AXIOM Aristos MX that will avoid these potential injuries in the future.
Initiated
2013-06-05
Root cause
Device Design
Terminated
2015-02-24
Distribution
Nationwide Distribution.
Recall number
Z-1795-2013
Recall status
Terminated
CDRH record ID
119698
FDA classified
2013-07-24
Recall event number
65604