Skip to content

Public FDA record

Z-1795-2020: CareFusion Safe-T-Centesis Catheter Drainage Tray, Sterile, For Single Use Only, Rx only, Catalog Number

Recall record Z-1795-2020 for Becton Dickinson & Company. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1795-2020
Company in FDA record
Becton Dickinson & Company
FDA record creation date
Apr 24, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The 8 Fr catheter was manufactured with the 6.9 Fr needle instead of the 8 Fr needle.

Source facts

Action
On March 26, 2020, an Urgent Medical Device Recall Notification was sent to consignees stating that Lot #0001304986 of the Safe-T-Centesis 8 Fr Catheter Drainage Trays shipped between June 27, 2019 and September 6, 2019 should be discarded. In addition, they were instructed to complete the Customer Response form regardless of whether or not they have any of the product. Customers with discarded product were instructed to indicate the number of affected products that were discarded on the response form and replacements would be issued to those customers.
Initiated
2020-03-26
Root cause
Mixed-up of materials/components
Terminated
2020-08-21
Distribution
Domestic distribution to the following states: Alabama, Colorado, Illinois, Massachusetts, Missouri, Oklahoma, Rhode Island, Virginia, and Wisconsin.
Recall number
Z-1795-2020
Recall status
Terminated
CDRH record ID
180725
FDA record created
2020-04-24
Recall event number
85380

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗