Open original FDA record ↗Recall record
- FDA identifier
- Z-1795-2020
- Company in FDA record
- Becton Dickinson & Company
- FDA record creation date
- Apr 24, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The 8 Fr catheter was manufactured with the 6.9 Fr needle instead of the 8 Fr needle.
Source facts
- Action
- On March 26, 2020, an Urgent Medical Device Recall Notification was sent to consignees stating that Lot #0001304986 of the Safe-T-Centesis 8 Fr Catheter Drainage Trays shipped between June 27, 2019 and September 6, 2019 should be discarded. In addition, they were instructed to complete the Customer Response form regardless of whether or not they have any of the product. Customers with discarded product were instructed to indicate the number of affected products that were discarded on the response form and replacements would be issued to those customers.
- Initiated
- 2020-03-26
- Root cause
- Mixed-up of materials/components
- Terminated
- 2020-08-21
- Distribution
- Domestic distribution to the following states: Alabama, Colorado, Illinois, Massachusetts, Missouri, Oklahoma, Rhode Island, Virginia, and Wisconsin.
- Recall number
- Z-1795-2020
- Recall status
- Terminated
- CDRH record ID
- 180725
- FDA record created
- 2020-04-24
- Recall event number
- 85380