Open original FDA record ↗Recall record
- FDA identifier
- Z-1796-2018
- Company in FDA record
- Siemens Healthcare Diagnostics, Inc.
- FDA record creation date
- May 10, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Calibration failure - confirmed a positive bias with LIPL lot FB8109 when compared to other Dimension LIPL lots. Lot FB8109 is not meeting Siemens internal specifications for lot-to-lot limits.
Source facts
- Action
- Customers are instructed to: - Discontinue use of Dimension LIPL Assay lot FB8109. - Recalibrate with an alternate lot of Dimension LIPL Assay. - Review their inventory of this lot to determine their laboratoryfs no-charge replacement needs. - Review the UMDR/UFSN letter with their Medical Director. - Complete and return the Field Correction Effectiveness Check Form attached to the UMDR/UFSN letter within 30 days. - If they have received any complaints of illness or adverse events associated with lot FB8109, immediately contact their local Siemens Customer Care Center or their local Siemens Technical Support Representative
- Initiated
- 2017-12-14
- Root cause
- Under Investigation by firm
- Terminated
- 2020-12-03
- Distribution
- Domestic and Foreign consignees
- Recall number
- Z-1796-2018
- Recall status
- Terminated
- CDRH record ID
- 162393
- FDA record created
- 2018-05-10
- Recall event number
- 79468