Z-1797-2020: Atellica CH 930 Analyzer, Siemens Material Number 11067000 with software version v1.21.0 SP1 or lower

Recall record Z-1797-2020 for Siemens Healthcare Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1797-2020
Company in FDA record
Siemens Healthcare Diagnostics, Inc.
FDA record creation date
Apr 24, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Editing an assay Test Definition and switching to another assay s Test Definition screen without saving the changes using the save button on the screen, may corrupt the settings of the assay s test definition, and Auto-Rerun or Auto-Dilution results may generate a result of Zero (0) for quantitative assays or Negative for qualitative assays if a well of the reagent pack is not calibrated.

Source facts

Action
An Urgent Medical Device Correction notification letter dated 2/27/20 was sent to customers. The purpose of this communication is to inform you of the issues listed in Table 2 below, that can occur on the Atellica CH 930 Analyzer, installed with Atellica Solution software (SW) versions v1.21.0 SP1 or lower and provide instructions on actions your laboratory must take. These issues will be corrected in SW v1.22.0 and v1.21.11, which will be available soon. Actions to be Taken by the Customer The following actions must be taken until your system has been updated to software version v1.22.0 (or V1.21.11 for Atellica Solution with decapper modules) or higher which resolves the issues listed above. 1. When making modifications to an assays Test Definition settings on the Setup>Test Definition>CH Test Definition screen, always follow the instructions outlined in the Atellica Solution online help 1.20 instructions Editing Assay Identification Parameters in CH Test Definition. Each Test Definition must be saved using the Save button on the CH Test Definition screen before starting modifications on another assays Test Definition. After changing any setting on the Definition, Calculation, or Calibration tabs of the CH Test Definition screen, select Save and then OK. Verify that QC results are not affected, and that results, units, and flags are reported correctly for any Test Definition that has been updated before running patient samples. 2. Ensure all reagent packs have a valid calibration (all wells) before running samples. This information is available on the screen Calibration > Calibration Overview. Refer to the Atellica Solution online help 1.20 instructions, Viewing Assay Reagent Calibration Order Status. When available, software v1.22.0 and/or v1.21.11 will be delivered as follows: For systems running software v1.19.0 or higher, that are connected to the Smart Remote Service (SRS), the software will be delivered electronically
Initiated
2020-02-27
Root cause
Under Investigation by firm
Terminated
2021-10-25
Distribution
US Nationwide. Global distribution.
Recall number
Z-1797-2020
Recall status
Terminated
CDRH record ID
180395
FDA record created
2020-04-24
Recall event number
85268

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗