Open original FDA record ↗Recall record
- FDA identifier
- Z-1798-2018
- Company in FDA record
- Inpeco S.A.
- FDA record creation date
- May 10, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Module may freeze without generating user warning. There is a potential risk in delay of sample processing, leading to delayed delivery of test results to patients.
Source facts
- Action
- On 2/13/18, distributors were notified by emailed Technical Service Bulletin. The recalling firm also released a Customer Letter to be forwarded to final customers (labs) by distributors. The Customer Letter advised users of the issue and informed that the recalling firm would be visiting sites to upgrade the firmware to resolve the issue. All impacted customers were instructed to complete and return the Receipt Confirmation within 7 days. Contact provided: Giorgia Amabile - Quality Assurance & Regulatory Affairs Manager, giorgia.amabile@inpeco.com, (+41) 91 9118 258.
- Initiated
- 2018-02-13
- Root cause
- Device Design
- Terminated
- 2019-04-03
- Distribution
- Distributors in 2 states: NY and IL.
- Recall number
- Z-1798-2018
- Recall status
- Terminated
- CDRH record ID
- 162888
- FDA record created
- 2018-05-10
- Recall event number
- 79645