Z-1798-2020: MoXy Laser Fiber, Material Number 0010-2400 - Product Usage: The fiber can be used for the surgical incis

Recall record Z-1798-2020 for Boston Scientific Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1798-2020
Company in FDA record
Boston Scientific Corporation
FDA record creation date
Apr 24, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

To reinforce existing instructions within the Directions for Use (DFU) and provide further guidance which will be added to the DFU that may reduce the likelihood of metal cap and fiber tip break temperature related complaints.

Source facts

Action
An Urgent Medical Device Product Advisory - Immediate Action Required dated 3/30/20 was sent to customers. Existing MoXy" laser fiber Direction for Use to be Reinforced 1- Ensure the distance from the fiber to the tissue is approximately 2mm (1 to 3 mm). Reference: Table 2. Troubleshooting and Procedural Considerations in product DFU. 2- Do not bury the tip of the fiber into the tissue. Reduced irrigation flow may result in damage to the fiber. Reference: Caution in page 3 of product DFU. 3- Failure to position the saline solution at least 106cm or not opening the flow valve will decrease the amount of fluid across the fiber end. Decreased fluid flow may limit the lasing time of the fiber. Reference: Caution in page 3 of product DFU. Section 8.3 of the MoXyTM Laser Fiber DFU has been updated to recommend an increased irrigation flow to increase the liquid cooling effect. The additions to the current DFU are highlighted in blue: 8.3 Performing the Procedure 1. Ensure user, patient, and personnel in the room are wearing appropriate protective eyewear before using the laser. WARNING: Do not press the footswitch without having the appropriate protective eyewear. Pressing the footswitch may activate the laser beam. Activation of the laser beam when not wearing protective eyewear may result in eye injury to the user, personnel in the room, or the patient. 2. Connect the tubing from the sterile saline for irrigation solution to the fiber luer lock connector. Position the saline solution at least 106 cm higher than the level of the endoscope/cystoscope. Increased irrigation flow by means of a saline pressure bag (set to 250 mmHg 300 mmHg) will further increase liquid cooling effect and may reduce fiber tip damage. Please complete and return the acknowledgement form to indicate receipt of this letter. Boston Scientific Quality Systems 763-494-1133 BSCFieldActionCenter@bsci.com
Initiated
2020-03-30
Root cause
Use error
Terminated
2023-04-05
Distribution
Worldwide distribution - US Nationwide and Global distribution.
Recall number
Z-1798-2020
Recall status
Terminated
CDRH record ID
180674
FDA record created
2020-04-24
Recall event number
85356

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗