Open original FDA record ↗Recall record
- FDA identifier
- Z-1798-2020
- Company in FDA record
- Boston Scientific Corporation
- FDA record creation date
- Apr 24, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
To reinforce existing instructions within the Directions for Use (DFU) and provide further guidance which will be added to the DFU that may reduce the likelihood of metal cap and fiber tip break temperature related complaints.
Source facts
- Action
- An Urgent Medical Device Product Advisory - Immediate Action Required dated 3/30/20 was sent to customers. Existing MoXy" laser fiber Direction for Use to be Reinforced 1- Ensure the distance from the fiber to the tissue is approximately 2mm (1 to 3 mm). Reference: Table 2. Troubleshooting and Procedural Considerations in product DFU. 2- Do not bury the tip of the fiber into the tissue. Reduced irrigation flow may result in damage to the fiber. Reference: Caution in page 3 of product DFU. 3- Failure to position the saline solution at least 106cm or not opening the flow valve will decrease the amount of fluid across the fiber end. Decreased fluid flow may limit the lasing time of the fiber. Reference: Caution in page 3 of product DFU. Section 8.3 of the MoXyTM Laser Fiber DFU has been updated to recommend an increased irrigation flow to increase the liquid cooling effect. The additions to the current DFU are highlighted in blue: 8.3 Performing the Procedure 1. Ensure user, patient, and personnel in the room are wearing appropriate protective eyewear before using the laser. WARNING: Do not press the footswitch without having the appropriate protective eyewear. Pressing the footswitch may activate the laser beam. Activation of the laser beam when not wearing protective eyewear may result in eye injury to the user, personnel in the room, or the patient. 2. Connect the tubing from the sterile saline for irrigation solution to the fiber luer lock connector. Position the saline solution at least 106 cm higher than the level of the endoscope/cystoscope. Increased irrigation flow by means of a saline pressure bag (set to 250 mmHg 300 mmHg) will further increase liquid cooling effect and may reduce fiber tip damage. Please complete and return the acknowledgement form to indicate receipt of this letter. Boston Scientific Quality Systems 763-494-1133 BSCFieldActionCenter@bsci.com
- Initiated
- 2020-03-30
- Root cause
- Use error
- Terminated
- 2023-04-05
- Distribution
- Worldwide distribution - US Nationwide and Global distribution.
- Recall number
- Z-1798-2020
- Recall status
- Terminated
- CDRH record ID
- 180674
- FDA record created
- 2020-04-24
- Recall event number
- 85356