Skip to content

Public FDA record

Z-1799-2015: Pull Reduction Device For 4.3mm Percutaneous Drill Guide; The pull reduction device with quick coupling i

Recall record Z-1799-2015 for Synthes, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1799-2015
Company in FDA record
Synthes, Inc.
FDA classification date
Jun 15, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Certain affected parts and lots of the Pull Reduction Devices may have been manufactured to an incorrect hardness specification which could result in intraoperative breakage.

Source facts

Action
DePuy Synthes sent an Urgent Notice Urgent notice dated May 28, 2015, to all affected customers to advise them of the defect and patient risks associated with its use. Customers were instructed if they have any of the identified devices, please take the following steps: Call DePuy Synthes at 1-800-479-6329 to obtain a Return Authorization (RA) Number. Complete the Verification Section (page 3 of this letter) by checking the appropriate box indicating affected product has been located. Also, please indicate the number of devices found and note the Return Authorization Number. Please include your name, title, address, telephone number and signature in the spaces provided. Return the Verification Section (page 3 of this letter) with the product to: Credit/Returns, DePuy Synthes, 1101 Synthes Avenue, Monument, CO 80132. Send a copy of the completed Verification Section by: Fax: (888) 943-5206 or Scan/email: Synthes2912@stericycle.com If you DO NOT have the identified product, please take the following steps: Complete the attached Verification Section (page 3 of this letter) by checking the appropriate box indicating that no affected product has been located. Please include your name, title, address, telephone number and signature in the spaces provided. This return documentation acknowledges your receipt of medical device removal information. Return the documents by: Fax: (888) 943-5206 or Scan/email: Synthes2912@stericycle.com Note: If the Verification Section is answered on behalf of more than one facility and/or individual, please clearly indicate the name and address of the facility and/or individual on page 3 of the notification. Customers with questions were instructed to call 610-719-5450. For questions regarding this recall call 610-719-5000.
Initiated
2015-05-28
Root cause
Component design/selection
Terminated
2016-08-22
Distribution
Worldwide Distribution-US (nationwide) and internationlly to Canada
Recall number
Z-1799-2015
Recall status
Terminated
CDRH record ID
137453
FDA classified
2015-06-15
Recall event number
71364

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗