Open original FDA record ↗Recall record
- FDA identifier
- Z-1800-2010
- Company in FDA record
- Inviro Medical Devices, Inc
- FDA classification date
- Jun 11, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Syringe Defects. Group 1 syringes: The plunger pull out force was below specification. Group 2 syringes: The needle length is out of specification.
Source facts
- Action
- Consignees were notified by letter on or about 05/10/2010. Consignees were instructed to review their inventory of InviroSnap Safety Syringes for the lot numbers and remove them from inventory and to return the affected product to Inviro Medical. They were instructed to pass on the information to anyone who uses or orders Inviro Snap Safety Syringes at their facility and to ensure a copy of the letter is provided to any other organizations to which affected devices have been transferred. A Response Card was included to be completed and faxed back to Inviro Medical Devices even if there is no product in inventory to verify receipt of notification. Questions regarding the recall were directed to the company's Customer Service number at (866) 914-2259. All product from Group 1 will be scrapped. Product from Group 2 will be retained in Quarantine until such time as the product can be exported the permission of the importing countries Ministry of Health.
- Initiated
- 2010-05-10
- Root cause
- Other
- Terminated
- 2011-12-08
- Distribution
- CA, FL, IL, MI, MO, MN, NC, OH, TX; and Canada
- Recall number
- Z-1800-2010
- Recall status
- Terminated
- CDRH record ID
- 91517
- FDA classified
- 2010-06-11
- Recall event number
- 55674