Z-1800-2025: Product Name: Ingenia 3.0T CX; Model Number: 781271;

Recall record Z-1800-2025 for Philips North America Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1800-2025
Company in FDA record
Philips North America Llc
FDA classification date
May 16, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for alignment errors in the cross reference line functionality when reviewing images generated with the MobiView application.

Source facts

Action
On May 12, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions that should be taken by the customer / user in order to prevent risks for patients or users You may continue to use your system(s) in accordance with the intended use. A. To avoid the potential issue of an incorrect cross reference line in MobiView generated images, use the same voxel size and field of view across all station scans. B. Do not use cross reference line functionality in box mode, all slices mode, or 3D mode. Use the cross reference line functionality in single slice mode. Circulate this notice to all users of this device so that they are aware of the issues and associated hazard/harm. Please retain this Urgent Medical Device Correction letter with your system(s) until the software upgrade is installed; ensure the notice is in a place likely to be seen/viewed. Please complete and return the attached response form to Philips MR promptly and no later than 30 days from receipt of this letter via email to: Philips.Recall@Philips.com. Completing this form confirms receipt of the Urgent Medical Device Correction letter, understanding of the issues, and required actions to be taken. 5. Actions planned by Philips MR to correct the problem A Philips representative will contact you to schedule time for a Field Service Engineer (FSE) to install a software upgrade to resolve the issues (reference FCO78100566, FCO78100584, FCO78100585, FCO78100620, FCO78100621). If you need any further information or support concerning this issue, please contact your local Philips representative. For North America, contact the Customer Care Solutions Center (1-800-722-9377, 8AM-8PM EST, Monday-Friday). This notice has been reported to the appropriate Regulatory Agencies. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, or by regular mail, or by fax. For more information, pl
Initiated
2025-05-13
Root cause
Software design
Distribution
Worldwide distribution - US Nationwide and the countries of Andorra, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, C¿te D'Ivoire, Cyprus, Czech Republic, Denmark, Finland, France, French Guiana, Georgia, Germany, Greece, Guadeloupe, Hong Kong, Iceland, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kenya, Latvia, Lebanon, Libya, Lithuania, Malaysia, Malta, Martinique, Mexico, Moldova, Monaco, Netherlands, New Zealand, Norway, Palestine, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., Serbia, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Kingdom, U.A.E.
Recall number
Z-1800-2025
Recall status
Open, Classified
CDRH record ID
213887
FDA classified
2025-05-16
Recall event number
96889

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗