Open original FDA record ↗Recall record
- FDA identifier
- Z-1801-2010
- Company in FDA record
- Ge Healthcare, Llc
- FDA classification date
- Jun 11, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
GE Healthcare has become aware of a safety issue associated with the CARESCAPE" Monitor B850 when monitoring ECG. The safety issue includes the potential for a delay in treatment.
Source facts
- Action
- The customers were sent on May 6, 2010, a GE Healthcare "Urgent Medical Device Correction" letter dated April 22, 2010. The letter was addressed to Healthcare Administrator / Risk Manager, Chief of Nursing, Director of Biomedical Engineering. The letter listed the Safety Issue, Affected Product Details, Product Correction, Safety Instructions and Contact Information. The customers were instructed to immediately discontinue use of the "Update Lead Set" function until the device software has been upgraded. If you have any questions regarding this medical device correction or identification of affected items, please contact Technical Support at 1-800-558-7044, or your local service representative.
- Initiated
- 2010-05-06
- Root cause
- Software design
- Terminated
- 2012-06-05
- Distribution
- Worldwide distribution: USA including CO, MA, NC, WI, and countries of UNITED KINGDOM, SWITZERLAND, SWEDEN, SPAIN, SINGAPORE, PORTUGAL, NORWAY NETHERLANDS, ITALY, INDIA, GERMANY, FRANCE, FINLAND, BELGIUM, and AUSTRALIA.
- Recall number
- Z-1801-2010
- Recall status
- Terminated
- CDRH record ID
- 90613
- FDA classified
- 2010-06-11
- Recall event number
- 55352