Open original FDA record ↗Recall record
- FDA identifier
- Z-1802-2018
- Company in FDA record
- Bard Peripheral Vascular Inc
- FDA classification date
- May 11, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Device is at risk of having detectable levels of bacterial endotoxin present on the packaging hoop that the catheter is packaged in. The bacterial endotoxin levels for the catheters are within acceptable levels; however, if bacterial endotoxins are present on the packaging hoops it may have the potential for cross-contamination in a clinical setting.
Source facts
- Action
- The firm initiated the recall via an Urgent Medical Device Recall Notification letter on 04/16/2018. The letter identified the affected device, reason for the recall, and required actions. Customers are asked to check their inventory for and remove the affected product code and lot number combinations. The Recall and Effectiveness Check Form should be completed and returned even if affected product has been used. The firm requested the return of the recalled product; and will provide a replacement after all information has been verified and product returned. The completed form can be emailed to BPV.CustomerSupportCenter@crbard.com or faxed to 1-800-994-6772.
- Initiated
- 2018-04-16
- Root cause
- Process control
- Terminated
- 2021-03-02
- Distribution
- Nationwide Distribution
- Recall number
- Z-1802-2018
- Recall status
- Terminated
- CDRH record ID
- 163824
- FDA classified
- 2018-05-11
- Recall event number
- 79896