Z-1804-2020: iLab Polaris Multi-Modality Guidance System; iLab Ultrasound Imaging System

Recall record Z-1804-2020 for Boston Scientific Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1804-2020
Company in FDA record
Boston Scientific Corporation
FDA record creation date
Apr 27, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Certain iLab Polaris Systems which were upgraded with a previous 2.10 software version (disk batch 629774) contained a bug in which only the True Color Secondary Capture (SC) format will archive Diastolic Hyperemia-Free Ratio data as intended. If either Multi-Frame UltraSound (US) or SC Image Storage formats are used, the DFR data will not be archived and the user will not be notified of this unexpected behavior.

Source facts

Action
On September 18, 2019, customers that were upgraded with version 54 of iLab Polaris 2.10 (disk batch 629774) software were sent a customer letter via overnight mail delivery. The letter provides instructions and recommendations to the user until a software upgrade (disk batch 636991) was completed. The firm informed that a firm representative will be in contact with customers directly to schedule a software upgrade at their facilities.The firm advises that systems can continue to be used during the time following the recommendations and instructions within the letter. Additionally, in the firm's Urgent Medical Device Correction notice it explains for Cath Lab Managers or Cath Lab Nurses to perform the following Actions: 1. Distribute the letter to system operators within their facility. 2. Immediately post a copy of Instructions to Archive DFR Data on or in a visible location near each system at their facility. The firm provides the following recommendations: If you possess a system with a serial number listed in the affected product listing note that: 1. After case closure and archiving the DFR data will be stored for only six (6) months on the system. After this timeframe the case will be automatically deleted from the system and cannot be retrieved. 2. The DFR data can be archived on CD/DVD, removable hard drive, or hospital network only by using the True Color SC format. You can ensure your data is archived using the True Color SC format following the "Instructions to Archive DRF Data". 3. Ensure that all cases are archived to prevent data loss. Note that archived cases will be automatically deleted after six (6) months. Additionally, the firm provides the following instructions: The Acknowledgement Form enclosed with the letter (Urgent Medical Device Correction) must be completed and returned even if you do not have any affected units. 1. Immediately post the information in a visible location near the product to ensure the in
Initiated
2019-09-18
Root cause
Other
Distribution
OUS: Japan, Belgium, Denmark, France, Great Britain, and Spain. No U.S. consignees.
Recall number
Z-1804-2020
Recall status
Completed
CDRH record ID
180401
FDA record created
2020-04-27
Recall event number
85271

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗