Z-1805-2026: Philips AneurysmFlow. Model Number: 001015. AneurysmFlow is a software medical device (Interventional T

Recall record Z-1805-2026 for PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1805-2026
Company in FDA record
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
FDA classification date
Apr 10, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential safety issue involving the Mean Aneurysm Flow Amplitude (MAFA) ratio of AneurysmFlow (Interventional Tool). Philips has determined that the MAFA ratio does not provide reliable prognostic information regarding aneurysm occlusion following Flow Diverter Stent treatment, however despite being stated in the IUF, it is being relied on for making clinical decisions.

Source facts

Action
Philips notified consignees on about 03/19/2026 via "URGENT Medical Device Correction" letter. Consignees were instructed NOT to use the MAFA ration for clinical decision making as indicated in the IFU and that AneurysmFlow may continue to be used in accordance with the Instructions for Use (IFU). Additionally, consignees were instructed to circulate the URGENT Medical Device Correction letter to all users so that they are aware of the issue and keep this URGENT Medical Device Correction letter with the documentation of the system until Philips corrects your system. Ensure the letter is in a place likely to be seen/viewed. Lastly, consignees were instructed to complete and return the response form. In case the Philips interventional X-ray system on which AneurysmFlow is installed has been transferred to another organization, consignees were requested to send a copy of the URGENT Medical Device Correction letter to that organization and inform Philips about this transfer through their local Philips representative. Philips is developing a software update to remove the MAFA ratio from AneurysmFlow. Philips expects to release this software update in December 2026. Local Philips representatives will contact consignees to schedule implementation of this software update once available.
Initiated
2026-03-19
Root cause
Labeling design
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Bnis insted Us Instas instes, Brazil, Canada, Chile, Colombia, Croatia, Czech Republic, Egypt, Estonia, Germany, Greece, Indonesia, Iraq, Italy, Japan, Kuwait, Latvia, Macao, Maldives, Mongolia, Morocco, New Zealand, Nicaragua, Poland, Portugal, Romania, Saudi Arabia, Serbia, Slovenia, South Africa, Sri Lanka, Switzerland, TURKEY, Ukraine, Utd.Arab.Emir., Uzbekistan, Vietnam.
Recall number
Z-1805-2026
Recall status
Open, Classified
CDRH record ID
219103
FDA classified
2026-04-10
Recall event number
98576

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗