Open original FDA record ↗Recall record
- FDA identifier
- Z-1806-2014
- Company in FDA record
- RAYSEARCH LABORATORIES AB
- FDA classification date
- Jun 17, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
This notice concerns a problem with the computation of ROI voxel volumes for cases where the CT has variable slice spacing. The dose grid volumetric representation of an ROI can be wrong. This affects all dose-volume properties for such ROIs, including DVH, dose statistics, clinical goals and constraints or objective functions. Furthermore, if material override ROIs are used, or if the external is not represented by contours, the bug can trigger an error in the 3D dose distribution.
Source facts
- Action
- On April 25, 2014, RaySearch emailed all consignees of the affected product a copy of the recall notice as well as the response forms. Back up notification methods include phone calls and site visits.
- Initiated
- 2014-04-25
- Root cause
- Software design
- Terminated
- 2017-06-26
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of AZ, CA, FL, IL, MA, MD, MO, MS, NJ, NV, NY, OH, PA, TN, VA, and WA, and the countries of Australia, Belgium, Germany, France, Italy, New Zealand, Republic of Korea, Switzerland, Spain, and United Kingdom.
- Recall number
- Z-1806-2014
- Recall status
- Terminated
- CDRH record ID
- 127076
- FDA classified
- 2014-06-17
- Recall event number
- 68162