Reason stated in the source
If the shaft breaks due to material fatigue, the trauma arm can fall down and may result in injury.
Source facts
- Action
- Philips issued Urgent Field Safety Notice (FCO72000015) sent 4/09/20 via certified mail states reason for recall health risk, action to take: The IfU (Instruction for Use) require the operator to not use the system and call for service, if a mechanical defect or malfunction is suspected. As soon as a deformation is determined, the system may no longer be used and call the service engineer. Should you feel uncertain regarding these instructions, please contact Philips. If you would like any further information or support concerning this issue, please contact your local Philips representative.
- Initiated
- 2020-04-07
- Root cause
- Device Design
- Terminated
- 2023-04-28
- Distribution
- Worldwide distribution - US Nationwide distribution including the states of CA, ME, RI, TX, WI and the countries of Canada, Australia, Austria, France, Germany, Greece, Hungary, Ireland, Netherlands, Portugal, Slovenia, United Kingdom.
- Recall number
- Z-1806-2020
- Recall status
- Terminated
- CDRH record ID
- 181038
- FDA record created
- 2020-04-28
- Recall event number
- 85472