- FDA identifier
- Z-1806-2025
- Company in FDA record
- Smith & Nephew, Inc.
- FDA classification date
- May 19, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Integrated Lag/Compression Screw Kit has a compression screw with hexagon socket that may be undersized, which causes the screw to be non-functional, which would require replacement with another adequate size screw, and exchange may exceed 30 minutes.
Source facts
- Action
- On 3/18/2025, recall notices were mailed and emailed to customers who were asked to do the following: 1) Inspect your inventory and locate any affected devices and quarantine them. 2) If you are a sales rep, district office, or distributor, you must notify your customers of the field action and ensure that the required actions are complete. 3) Complete and return the notice response section via email to FieldActions@smith-nephew.com If you have any questions email the firm at the email address above.
- Initiated
- 2025-03-18
- Root cause
- Process control
- Distribution
- Worldwide - US Nationwide distribution in the states of FL, NH, TX, CA, OH, NY, AL, IA, MA, NV, MO, WA, LA, TN, NC, KY, NM, NJ, AZ and the country of ZA.
- Recall number
- Z-1806-2025
- Recall status
- Open, Classified
- CDRH record ID
- 213212
- FDA classified
- 2025-05-19
- Recall event number
- 96586