Z-1808-2009: American Optisurgical, Horizon Phacoemulsification System, Model: HZN-2030, for ophthalmic surgery.

Recall record Z-1808-2009 for American Optisurgical Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1808-2009
Company in FDA record
American Optisurgical Inc
FDA classification date
Sep 4, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Device operation interrupted: A software bug causes Phaco to stop when vacuum goes above 300 mmHg while in Multi Burst mode. Use of this product at the parameters specified above could lead to a delay in surgery that might present a health risk.

Source facts

Action
American Optisurgery sent an E-mail to customers on April 2, 2009 stating the reason for recall and informed them to immediately examine their inventory and quarantine product subject to recall. In addition, if they have further distributed this product, please identify customers and identify them at once of this product recall. Their notification to customers may be enhanced by including a copy of this recall notification letter. They were then told to refer to the enclosed Recall Questionnaire for instructions on what to do with the recalled product: The recall should be carried out to the user level and return the enclosed response form as soon as possible. If there are any questions, they are advised to contact David Salzberg at (800) 576-1266.
Initiated
2009-03-24
Root cause
Software change control
Terminated
2009-09-04
Distribution
Worldwide distribution: USA (California), Brazil, Algeria, France, South Korea, India, and Peru.
Recall number
Z-1808-2009
Recall status
Terminated
CDRH record ID
83671
FDA classified
2009-09-04
Recall event number
52697

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗