Skip to content

Public FDA record

Z-1809-2024: LASER Application Instrument, 23 cm, REF: 8694; LEUNIG-GREVERS LASER Application, REF: 461000

Recall record Z-1809-2024 for Karl Storz Endoscopy. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1809-2024
Company in FDA record
Karl Storz Endoscopy
FDA classification date
May 13, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Laser surgical instruments have no specific evidence of a validated reprocessing method, therefore these instruments lack sterility assurance, so once the instruments have gone through reprocessing after use and then are used again, there is an increased risk of the patient being exposed to infection.

Source facts

Action
On 4/1/24, recall notices were mailed to representatives for medical product safety, users, operators, distributors who were asked to do the following: 1) Quarantine and discontinue use of affected devices. 2) Pass on the recall notice to all users of the affected devices and to all other persons who need to be aware within your organization. 3) If you have or may have distributed the products listed, please identify and promptly notify those recipients, or provide the recalling firm a list of customers who received/may have received the products listed. 4) Complete and return the customer reply form via email to karlstorz4835@sedgwick.com 5) If you have product to return contact the firm at the following phone numbers: For Human Product Telephone: 800-421-0837 Option 1 For Vet Product Telephone: 800-955-7832 Option 1 The firm has placed all affected product on hold and has ceased the distribution of the affected product in the U.S. until further notice. For all other questions regarding this communication, please send an email to KSEAProductLabeling@karlstorz.com
Initiated
2024-04-01
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide in the states of OH, CA, FL, MA, TX, KY, WI, MI, MN, AR, NE, IL, NY, HI and the country of Argentina.
Recall number
Z-1809-2024
Recall status
Open, Classified
CDRH record ID
207145
FDA classified
2024-05-13
Recall event number
94389

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.