Lupine BR Anchor w/Orthocord was assembled with an incorrect suture configuration. The suture loop was not assembled at the distal end of the anchor. It was only threaded through the suture hole in the anchor. Risk associated with incorrect suture configuration include premature breakage.
Source facts
Action
DePuy Mitek notified customers by letter dated April 29, 2008 titled Urgent: Voluntary Medical Device Recall Letter asking consignees to isolate and do not use product and to return any of the identified product to DePuy Mitek using enclosed prepaid Federal Express shipping labels and to include a photocopy of the completed Business Reply Form. The firm is asking customers to complete the Business Reply Form with a self addressed, stamped envelope indicating level of inventory of the affected product and to return for immediate delivery of replacement product. If you have any questions contact DePuy Mitek Customer Service directly at 1-800-382-4682.
Initiated
2008-04-29
Root cause
Employee error
Terminated
2008-10-02
Distribution
Class II Recall - Worldwide Distribution --- USA including states of CA, NY, MA, NV, OH, TX, TN, MD, and countries of France, UK, Canada, Spain.
Recall number
Z-1810-2008
Recall status
Terminated
CDRH record ID
70482
FDA classified
2008-09-17
Recall event number
47897