Skip to content

Public FDA record

Z-1810-2008: Lupine BR Anchor w/Orthocord TCP/PLGA Absorbable Anchor, size 2 (5 metric) Braided Composite Suture, viol

Recall record Z-1810-2008 for DePuy Mitek, Inc., a Johnson & Johnson Co.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1810-2008
Company in FDA record
DePuy Mitek, Inc., a Johnson & Johnson Co.
FDA classification date
Sep 17, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Lupine BR Anchor w/Orthocord was assembled with an incorrect suture configuration. The suture loop was not assembled at the distal end of the anchor. It was only threaded through the suture hole in the anchor. Risk associated with incorrect suture configuration include premature breakage.

Source facts

Action
DePuy Mitek notified customers by letter dated April 29, 2008 titled Urgent: Voluntary Medical Device Recall Letter asking consignees to isolate and do not use product and to return any of the identified product to DePuy Mitek using enclosed prepaid Federal Express shipping labels and to include a photocopy of the completed Business Reply Form. The firm is asking customers to complete the Business Reply Form with a self addressed, stamped envelope indicating level of inventory of the affected product and to return for immediate delivery of replacement product. If you have any questions contact DePuy Mitek Customer Service directly at 1-800-382-4682.
Initiated
2008-04-29
Root cause
Employee error
Terminated
2008-10-02
Distribution
Class II Recall - Worldwide Distribution --- USA including states of CA, NY, MA, NV, OH, TX, TN, MD, and countries of France, UK, Canada, Spain.
Recall number
Z-1810-2008
Recall status
Terminated
CDRH record ID
70482
FDA classified
2008-09-17
Recall event number
47897

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗