Skip to content

Public FDA record

Z-1810-2009: CyberKnife Treatment Delivery System, a subsystem of the CyberKnife Robotic Radiosurgery System, manufact

Recall record Z-1810-2009 for Accuray Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1810-2009
Company in FDA record
Accuray Inc
FDA classification date
Sep 14, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Targeting accuracy out of specification, Error alert does not render the system down, which may result in mistreatment in the wrong area.

Source facts

Action
Accuray, Inc. issued an "Urgent Device Correction" notice dated June 22, 2009. The consignees were informed of the defect including methods for detecting the error. The firm is in the process of developing a software patch to address the issue. For further information, contact Accuray, Inc. Customer Support at +1-877-668-8667 (US) or +1-408-716-4700 (outside US) or customersupport@accuray.com.
Initiated
2009-06-22
Root cause
Software design
Terminated
2011-01-05
Distribution
Worldwide Distribution -- US including states of MA, FL, AL, TN, DE, CA, NJ, OK, CO, NC, IL, PA, MD, DC, MS, CN, NV, RI, SC, UT, VA, MI, NE, MO, TX and AL and countries of Canada, Russia, UK, Ukraine, Saudi Arabia, Switzerland, Turkey, Taiwan, India, China, Korea and Thailand.
Recall number
Z-1810-2009
Recall status
Terminated
CDRH record ID
83540
FDA classified
2009-09-14
Recall event number
52643

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗