Open original FDA record ↗Recall record
- FDA identifier
- Z-1810-2009
- Company in FDA record
- Accuray Inc
- FDA classification date
- Sep 14, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Targeting accuracy out of specification, Error alert does not render the system down, which may result in mistreatment in the wrong area.
Source facts
- Action
- Accuray, Inc. issued an "Urgent Device Correction" notice dated June 22, 2009. The consignees were informed of the defect including methods for detecting the error. The firm is in the process of developing a software patch to address the issue. For further information, contact Accuray, Inc. Customer Support at +1-877-668-8667 (US) or +1-408-716-4700 (outside US) or customersupport@accuray.com.
- Initiated
- 2009-06-22
- Root cause
- Software design
- Terminated
- 2011-01-05
- Distribution
- Worldwide Distribution -- US including states of MA, FL, AL, TN, DE, CA, NJ, OK, CO, NC, IL, PA, MD, DC, MS, CN, NV, RI, SC, UT, VA, MI, NE, MO, TX and AL and countries of Canada, Russia, UK, Ukraine, Saudi Arabia, Switzerland, Turkey, Taiwan, India, China, Korea and Thailand.
- Recall number
- Z-1810-2009
- Recall status
- Terminated
- CDRH record ID
- 83540
- FDA classified
- 2009-09-14
- Recall event number
- 52643