Skip to content

Public FDA record

Z-1810-2010: Cardiac Science Powerheart 9390A automated external defibrillator. This fully automatic model does not r

Recall record Z-1810-2010 for Cardiac Science Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1810-2010
Company in FDA record
Cardiac Science Corporation
FDA classification date
Jun 21, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The K302 relay switch of the G3 series AED product line may experience an early failure. If the K302 contacts are open at the time of rescue, there is a possible delay in therapy or failure to deliver therapy.

Source facts

Action
Customer notification letters were mailed to US end users and distributors (final consignees) between December 18 and 23, 2009 by USPS certified mail, return receipt requested. Notification letters were sent by FedEx to Canadian distributors (final consignees) on December 23, 2009. Secondary consignee (distributor) letters were mailed on December 23, 2009. Letters advise user to check the status indicator on front of AED to determine if it is rescue ready. Users are advised to keep the AED in service and to perform the Daily and Monthly Scheduled Maintenance as outlined in the Operator and Service Manual. Questions about this communication should be addressed to 1.877.901.1788 toll free from inside the US, or to email .
Initiated
2009-12-18
Root cause
Other
Terminated
2011-12-08
Distribution
Domestic: 323 units to distributors and end users. Canada: 9 units to 2 distributors.
Recall number
Z-1810-2010
Recall status
Terminated
CDRH record ID
91319
FDA classified
2010-06-21
Recall event number
55603

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗