Open original FDA record ↗Recall record
- FDA identifier
- Z-1811-2015
- Company in FDA record
- CareFusion 303, Inc.
- FDA classification date
- Jun 16, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
An issue with the cancel functionality used during atypical infusion programming to cancel user inputted values. An infusion may start that is greater than or less than the hospital established limits for the specific medication.
Source facts
- Action
- CareFusion sent an Urgent Medical Device Recall Notification letter dated May 13, 2015 to customers. The letter identified the affected product, problem and the recommended actions to be taken. Customers are informed that CareFusion has released a software upgrade to address this issue. CareFusion will contact all affected customers within 60 days to initiate the scheduling process for the software upgrade installation. Customers with recall related questions are instructed to contact CareFusion Support Center at (888) 562-6018, 7am-4pm PT, supportcenter@carefusion.com. Customers with adverse event report questions are instructed to contact customer advocacy at (888) 812-3266, 24 hrs/day, 7 days a week, customerfeedback@carefusion.com. Customers with technical questions regarding the Alaris System are instructed to contact Technical Support at (888) 812-3229, 6am-5pm PT, Mon-Fri, dl-us-inf-techsupport@carefusion.com.
- Initiated
- 2015-05-13
- Root cause
- Software design
- Terminated
- 2016-05-20
- Distribution
- US Nationwide Distribution
- Recall number
- Z-1811-2015
- Recall status
- Terminated
- CDRH record ID
- 136048
- FDA classified
- 2015-06-16
- Recall event number
- 71081