Skip to content

Public FDA record

Z-1811-2024: Optum Nimbus II Plus, Model number AAA-00170; Ambulatory Infusion Pump

Recall record Z-1811-2024 for OptumHealth Care Solutions LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1811-2024
Company in FDA record
OptumHealth Care Solutions LLC
FDA classification date
May 21, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Nimbus II Plus, Infusion Pump Systems, manufactured by InfuTronix, are being removed the from the market due to multiple potential failure modes, including battery failure, upstream occlusion, system errors, drug product leakage, high or low flow rate, or damaged pump housing.

Source facts

Action
Optumhealth Care Solutions, LLC issued an Urgent Device Recall by the Manufacturer of Nimbus II Pump notice to its consignees on 04/18/2024 via email Delivery/Adobe Sign and UPS Next Day Air. The notice sent an electronic link to the press release and stated that a nurse would be contacting the consignee to schedule a pump replacement, including a home visit to instruct the patient on the use of the new pump. In the meantime, if you have any issue with your current Nimbus II pump, please contact your nurse right away to prioritize your pump replacement. Additionally, the manufacturer recommends in its recall notice that you take the following actions, so please be aware and comply with them until your Nimbus II pump has been replaced: 1. When replacing batteries, only new batteries should be used. Care should be taken to ensure that old and new batteries are not confused when completing the replacement. 2. The battery should not be cycled (turned on/off) to clear the occlusion alarm. If the occlusion alarm cannot be cleared, a new pump should be used. Silencing the alarm does not clear the occlusion. 3. Due to the potential ambulatory use of the pump, you may carry the drug product dispensing pouch in a carrying pack. Please note that impeding the tubing set or placing pressure on the device (leaning against it, sitting on it) may cause the device to malfunction. 4. If you note any difficulties or abnormalities with the pump or accessories, immediately discontinue use of the pump and notify your Optum nurse or health care provider.
Initiated
2024-04-18
Root cause
Device Design
Distribution
US Nationwide Distribution
Recall number
Z-1811-2024
Recall status
Open, Classified
CDRH record ID
207307
FDA classified
2024-05-21
Recall event number
94491

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗