Open original FDA record ↗Recall record
- FDA identifier
- Z-1811-2025
- Company in FDA record
- MEDLINE INDUSTRIES, LP - Northfield
- FDA classification date
- May 31, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Affected kits contain recalled Integra Lifesciences Codman Disposable Perforators due to inadequate welds on specific lots. Perforators with inadequate welds may break or separate.
Source facts
- Action
- On April 22, 2025, the firm notified customers of the recall through letters titled Immediate Action Required Voluntary Medical Device Recall. Customers were instructed to identify and quarantine all affected product on hand. Customers are to submit a response form to the recalling firm. Upon receipt of the response firm, Medline will provide the customer with over-labels to place on affected inventory, which will include instructions for staff to remove and discard the affected component prior to using the kit. Distributors should notify downstream customers/accounts of this recall. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
- Initiated
- 2025-04-22
- Root cause
- Nonconforming Material/Component
- Distribution
- US distribution to CA, FL, IL, IN, KY, MA, MD, MN, NY, TX, VT, WA and WI. No OUS distribution.
- Recall number
- Z-1811-2025
- Recall status
- Open, Classified
- CDRH record ID
- 213732
- FDA classified
- 2025-05-31
- Recall event number
- 96821