Open original FDA record ↗Recall record
- FDA identifier
- Z-1812-2015
- Company in FDA record
- Intuitive Surgical, Inc.
- FDA classification date
- Jun 17, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The coating of the 5 mm Thoracic Grasper main tube may experience damage with the possibility of fragments of coating material dislodging and potentially falling into patient.
Source facts
- Action
- Urgent Medical Device Recall letters were sent to all consignees on May 21, 2015. The letter describes the affected product and issue involved. Additionally, the risk to health, as well as the actions to be taken were also discussed in the letter. Customers are to ensure that all affected personnel were informed of the notice. All affected product are to be identified and returned by contacting Customer Service for a RMA. The Acknowledgment Form is to be completed and returned using the instructions provided. If further information or support concerning this recall is needed, customers are to contact their local Clinical Sales Representative or Intuitive Surgical Customer Service.
- Initiated
- 2015-05-21
- Root cause
- Nonconforming Material/Component
- Terminated
- 2018-07-02
- Distribution
- Worldwide Distribution -- US, Belgium, Brazil, Canada, Chile, Colombia,, Denmark, France, Germany, India, Netherlands, Panama, Singapore, South Korea, Switzerland, Taiwan, Turkey, and United Kingdom.
- Recall number
- Z-1812-2015
- Recall status
- Terminated
- CDRH record ID
- 137239
- FDA classified
- 2015-06-17
- Recall event number
- 71285