Operators noticed unwanted release of evaporating anesthetic agent at the end of the filling procedure. Eye contact with volatile anesthetics can result in serious irritation. Immediate eye flushing and consultation of a physician is recommended. Updated Material Safety Data Sheets (MSDS) for Desflurane recommend wearing safety glasses when handling these substances.
Source facts
Action
Drager mailed a "Medical Device Safety Alert" dated April 21, 2016, to customers informing them of the problem and provided instructions for the "correct filling procedure" and recommendations on how to prevent evaporation and spraying of Desflurane after bottle release to resolve the "High Friction of gas bottle valve" problem.
Initiated
2016-04-21
Root cause
Process change control
Terminated
2016-12-07
Distribution
US nationwide distribution, including Puerto Rico and Guam.
Recall number
Z-1812-2016
Recall status
Terminated
CDRH record ID
145674
FDA record created
2016-05-25
Recall event number
73959